CPET - Automated Input Measurement
15O Arterial Time Activity Curve (TAC)
Questions about this project should be addressed to: Brian Murphy - brian@petnet.buffalo.edu.
Purpose
Develop an automated system for measuring arterial Time-Activity curves (TAC) during
15O-water studies. These curves are required in order to properly compute quantitative
Cerebral Blood Flow using the autoradiographic technique (and most other techniques as
well).
Prior to the development of this system, TACs were obtained manually by withdrawing
arterial blood into weighed test tubes. These tubes were subsequently counted in an
automated counting device to determine the activity in each sample. This approach has
several significant drawbacks:
- sampling rate is low; maximum rate is one sample every 3-5 [sec]. This results in a
large amount of dispersion (and poor sampling resolution) in the resulting TAC which in
turn introduces errors into the calculated CBF values.
- sample withdrawal rate is variable (depends on varying arterial pressure).
- large amount of manual sample processing resulting in
- significant complication of the study during the acquisition of the samples (requiring
1-2 people focused on that task alone)
- significant amount of 15O decay between sample acquisition and sample counting
- increased personnel radiation exposure
With the new system, blood is automatically withdrawn at a uniform rate (5-20[ml/min])
and almost instantaneously counted in much shorter sampling intervals (0.5[sec]). This
eliminates all of the drawbacks present in the manual sampling method.
The beta detector approach was chosen because the high 511[keV] photon background
levels from 15O present in the patient make TAC determination by measuring these photons
difficult without a large amount of shielding. This shielding, in turn, would necessitate
moving the detector assembly further from the patient - thus increasing the length of
tubing through which the sample is acquired and adding dispersion to the resulting TAC.
Coincidence detection is another approach, but was deemed unnecessary for this project -
in large part due to increased detector/electronics complexity and the high cost (5 to 7
times that of current system costs) associated with commercial coincidence systems. With
the beta detector, we are able to place the detector assembly only centimeters away from
the arterial line catheter.
Several papers have been previously published on the use of a beta detector in this
capacity. The following paper is a good place to start: A System for Cerebral Blood
Flow Measurement Using an H215O Autoradiographic Method and Positron Emission
Tomography I.Kanno, et. al.; J of Cerebral Blood Flow and Metabolism; Vol 7, No
2, 1987; pg. 143-153.
Thanks to:
- Joe Villani, University at Buffalo Dept. of Nuclear Medicine - for supplying the Beta
Detector, some electronics/cables, and the shielding
- Alan Lockwood, University at Buffalo Dept. Nuclear Medicine and VAMC - for equipment
funding
- Ed Bednarczyk, University at Buffalo Dept. Nuclear Medicine - for tubing, plumbing
pieces parts and discussion of system he used in Cleveland.
- Raymond Raylman, University of Michigan Ann Arbor - for discussion of UM's beta detector
design
Equipment
- Electronics
- 486-33 SX acquisition computer w/ ISA card slot for MCS-plus(TM)
- EG&G Ortec ACEMate(TM) Model 925-SCINT Amplifier and Bias Supply
- EG&G Ortec MCS-plus(TM)
- SCA sweep mode (10V / 128 channels) to emulate MCA acquisition
- MCS sweep mode with independently adjustable lower and upper discriminator window
thresholds
- PMT base to signal/HV splitter and signal/HV cables (courtesy Joe Villani)
- Ludlum Model 44-1 Beta Survey Detector (courtesy Joe Villani)
- 0.25[cm] thick x 11.6[cm^2] surface area window (plastic scintillator)
- 0.8 [mg/cm^2] metalized mylar over detector window for light tightness
- 2[in] OD x 6.5[in] length self contained 10 stage PMT/detector package
- Shielding
- Octagonal machined lead casing with handle surrounding PMT/detector assembly;
25.5x10x10[cm] outside dimensions, 2.54[cm] thick (courtesy Joe Villani)
- Harvard '33' Syringe Pump
- catheter assembly
- detector tubing: Teflon coated; OD=0.5[mm], ID=0.41[mm] (courtesy Ed Bednarczyk)
Operating Design/Characteristics
- tubing
- detector to patient tubing length (excluding catheter): 20[cm]. This may be reduced in
the future (7-10 [cm] is feasible)
- tubing is coiled (~0.1[ml]) in front of detector; an additional ~0.1[ml] of entry/exit
tubing that will contribute to some of the counts seen by the system.
- efficiency*geometry*volume correction factor for converting to actual [µCi/ml]
from [µCi] measured by the detector was empirically determined to be 51.1. Given that the
detector sees only 0.1-0.2[ml] of sample at any given time and is only looking at roughly
50% of the positrons escaping the tubing, this translates to an approximate beta detection
efficiency of 18-36%. The detector has a quoted efficiency of 15% for 14C and 60% for 32P.
- Dispersion in the detector system: External dispersion of the input system (e.g. that
occurring outside the body and due to the catheter/detector assembly) can be determined by
measuring the rising/falling edges of the detector system's response to an input step
function (Error Analysis of a Quantitative Cerebral Blood Flow Measurement Using H2-15O
Autoradiography and Positron Emission Tomography, With Respect to the Dispersion of the
Input Function, H. Iida, et.al., J of Cerebral Blood Flow and Metabolism; Vol
6, No 5, 1986; pg. 536-545). Using the data from the tail portion of the subject raw TAC below, one can see that the time to rise or
fall to half-maximum in the tail section of the curve when the 3-way stopcock was being
switched between heparinized saline and the arterial line is always <= 1[sec].
Approximating the rising/falling curve as a mono-exponential of the form f(t)=exp(-t/tau)
to gives us a tau of < (1.0[sec]/-ln(0.5)) or tau < 1.5[sec], which can be
considered negligible.
- Linear activity range: Approximately 30 to < 0.5[µCi/ml].
- Background: measurements show that the background contribution of 511[keV] photons
doesn't exceed 0.15[µCi/ml] in a typical 15-O water study with an initial injection of
70[mCi].
Example Data Curves
Representative TAC
These curves show the raw TAC, a quantitative TAC derived from the raw TAC, and the
typical background underlying the quantitative TAC.
- subject raw TAC (14,412
bytes) This image shows the raw Time Activity Curve (TAC) as it appears on the MCS
acquisition display. It was acquired during a ~10[sec] slow bolus injection of 70[mCi] of
15O -water. Sampling was begun at the start of injection and was performed at 0.5[sec]
intervals for 3[min] at a withdrawal rate of 12[ml/min]. Initial activity onset to initial
curve peak (indicated by vertical line, counts in this channel listed at bottom of
display) occurs over 18[sec]. In the >120[sec] section of the TAC, the input to the
detector was switched several times via a 3-way stopcock between the arterial line and a
pressurized IV line filled with heparinized saline (with the withdrawal pump running
continuously). This permits us to measure the 511[keV] gamma background and provides the
raw data necessary for us to determine dispersion in the detector system.
- subject quantitative TAC (6,219 bytes) This plot shows
the same TAC as above in [µCi/cc]/[sec] for the first 120[sec] after a) decay correction
of 15-O for count duration and back to time of injection, b) correction for volume of 15-O
in front of the detector, and c) correction for detector geometry/efficiency factors.
- subject quantitative background (7,129 bytes)This plot
shows the background counts seen by the detector system over the first 120[sec] (same
conditions as above). It was acquired by flushing the detector tubing with heparinized
saline and counting as above following the subject's 2nd 15-O injection (no withdrawal
pump used and no arterial blood in front of the detector). You can see from the plot that
this background never exceeds a contribution of ~0.15[µCi/cc] to the overall TAC.
- hand sampled quantitative TAC (5,465 bytes) This plot
shows a TAC acquired by the old hand-sampling technique in [nCi/cc]/[sec]. In particular,
note the severe undersampling on the curve's peak (basically 4 points covering the entire
rise and initial fall).
15O Calibration
The data in the following curves were derived by drawing 1.0[mCi/cc] of 15O -water into
the detector assembly and initiating counting measurements after the activity had decayed
to a level at the high end of the range normally observed in patient TAC curves (in this
case, we waited 4 x 123[sec] half-lives). Data was then acquired for 6 half-lives to
determine the response characteristics of the detector system over the entire range of TAC
levels expected.
- raw data (18,689 bytes) As it appears on the MCS
acquisition screen. The main graph is integrated counts over the 0.5[sec] counting
interval for each channel. The small graph in the upper right hand corner is this same
graph with ln() scaling on the Y-axis.
- [µCi/ml] (5,218 bytes) Data after applying analytical
corrections for decay during the 500[ms] counting interval and empirically determined
correction factors for volume of sample in front of the detector, geometry of
sample/detector, and efficiency of detector. The actual data between the two vertical
lines (bounded with triangles) is fit to determine the half-life of the sample and its
actual activity in [µCi/ml]. In this case, the half-life was found to be 123.099[sec]
which is in good agreement with the known half-life of 15O. An ideal curve is then
generated using this data and overlaid on the raw data to illustrate count rates where
dead time may pose a problem (e.g. where ideal curve and actual data deviate).
- ln([µCi/ml]) (5,367 bytes) Same as [µCi/ml] graph, but
after taking the natural log of the activity
- Percent Error (19,347 bytes) [µCi/ml] fit vs. measured.
This graph plots the percent error of theoretical (fit) data vs. actual (measured) data
based on the two curves plotted in the [µCi/ml] plot above.
Sensitivity to 511[keV] background from patient
The data in the following curves are data as acquired from a study where the subject
was injected with 70[mCi] of 15O -water. The "Marker Channel" is the current
channel who's data is displayed at the bottom of the screen. This channel is represented
visually on the display by a vertical white line running the full vertical width of the
plot.
- signal & background (14,915 bytes) Tail end of TAC
acquisition (from approximately 120-180[sec] post injection of quantitative TAC above)
where, at the end of the acquisition, the detector line was switched three times between
the arterial line and a heparinized saline IV line (third switch is at last few points of
the acquisition). You can see the rapid drop in counts which occurs when the saline
reaches the detector. At this point, until the line is switched back to the arterial line,
all counts (or lack thereof) are coming from background 511[keV] photons.
- background (14,347 bytes) Acquisition of background data
only (from 0-120[sec] post injection, no blood drawn through the detector system following
injection). Compare the maximum count rate seen here ~126[cps] with ~17,660[cps] seen at
the 20[µCi/ml] level and ~800[cps] seen at the 1[µCi/ml] level.
- SCA sweep (11,075 bytes) This image shows an SCA sweep
from our detector system while measuring background from a subject (no activity in the
detector's tubing). In Single Channel Analyzer (SCA) sweep mode, a very narrow
acceptance window (a "single channel") is rapidly and repeatedly swept across
the entire voltage range (0-10[V]) of the detector's amplified output, dividing that range
into 128 separate channels. Events detected which result in a particular output voltage
are then accumulated in the appropriate channel. Note that this is the "poor
man's" equivalent of a Multi-Channel Analyzer (MCA) acquisition (where all
channels are observed simultaneously) since SCA sweep only records an event at a
particular voltage if that event's voltage matches the current SCA sweep channel. Our
hardware does not support MCA acquisitions, however, the SCA sweep is sufficient for
acquiring an energy spectrum of incoming events from a relatively stable source. The
position and width of the voltage acceptance window utilized in Multi-Channel Scalar
(MCS) mode, which records the number of events falling within the voltage acceptance
window over multiple time points during TAC acquisition, can therefore be determined and
set so that background and other signals we aren't interested in can be omitted from the
TAC. For our studies, the window is set to exclude the voltages at the low end of the
spectrum (more easily seen as the larger peak at the leftmost edge of the inlaid semi-log
plot), and to include everything above that up to the 10[V] limit.
Other sites and their solutions
coincidence detector - R. Paul
Maguire; PET Program, Paul Scherrer Institute
Future Work
- possibly incorporate dead time corrections to permit measurement of higher [µCi/ml]
levels
- link detector system, Harvard pump, and PET camera so that acquisition start may be
accomplished with a single button press
- incorporate 15O bolus infusion via Harvard pump
- see if we can see any of the lower-energy betas produced by 18F, and see if
signal/background is high enough to make this device useful for 18FDG TAC determinations.
- See if we can develop a non-invasive system for capturing 15O
input curves.